A Comparative Evaluation of Stress Distribution in Different Designs of Corticobasal Implants in Bone

A 3-Dimensional Finite Element Analysis

Autori

  • Sneha Gontu PG Student, Department of Prosthodontics and Crown & Bridge, Sree Sai Dental College and Research Institute, Srikakulam, Andhra Pradesh, 532001, India
  • Ch Siddesh Kumar MDS, Professor & Head of the Department, Department of Prosthodontics and Crown & Bridge, Sree Sai Dental College and Research Institute, Srikakulam, Andhra Pradesh, 532001, India
  • Susmita Mondal PG Student, Department of Prosthodontics and Crown & Bridge, Sree Sai Dental College and Research Institute, Srikakulam, Andhra Pradesh, 532001, India
  • Jwalithaclare Mesa PG Student, Department of Prosthodontics and Crown & Bridge, Sree Sai Dental College and Research Institute, Srikakulam, Andhra Pradesh, 532001, India
  • L. Srikanth MDS, Reader, Department of Prosthodontics and Crown & Bridge, Sree Sai Dental College and Research Institute, Srikakulam, Andhra Pradesh, 532001, India
  • Bhavana Tatineni PG Student, Department of Prosthodontics and Crown & Bridge, Sree Sai Dental College and Research Institute, Srikakulam, Andhra Pradesh, 532001, India

Cuvinte cheie:

Basal implants, KOS implants, KOS Plus implants, Finite element analysis, Axial and non-axial load

Rezumat

Purpose: The purpose of this study is to use a three-dimensional finite element analysis to assess the stress distribution around bone for different corticobasal implant designs.

Materials and Methods: A three-dimensional model of a human mandibular bone was created using Finite Element Analysis (FEA) modeling software, based on measurements from a dried human edentulous mandible. This model was duplicated to evaluate stress distribution around the bone with different implant designs: KOS Implant (rough surface), KOS Implant (smooth and rough surface), KOS PLUS Implant (smooth and rough surface), and KOS MU Implant (rough surface). Each implant measured 12 mm in length and 3.7 mm in width. Axial and non-axial loads of 200 N and 20 N, respectively, were applied to the implant in these FEA models. The models were then analyzed using ANSYS software (version R18.1) for finite element analysis.

Results: For the group I implant models subjected to an axial load of 200 N, the overall stress (MPa) values for the KOS Classic implants were 257.515 MPa. Under a non-axial load of 20 N applied to the group II implant models, the KOS Classic implants exhibited overall stress values of 142.424 MPa.

Conclusion: This finite element comparative study found that the KOS Classic implant models exhibited the maximum Von Mises Stress distribution. Further research is necessary to comprehend the clinical implications of these findings.

##plugins.themes.default.displayStats.downloads##

##plugins.themes.default.displayStats.noStats##

Publicat

2024-10-08

Număr

Secțiune

Articles

Cum cităm

Gontu, S., Kumar, C. S., Mondal, S., Mesa, J., Srikanth, L., & Tatineni, B. (2024). A Comparative Evaluation of Stress Distribution in Different Designs of Corticobasal Implants in Bone: A 3-Dimensional Finite Element Analysis. TWIST, 19(4), 63-70. https://twistjournal.net/twist/article/view/513

##plugins.generic.shariff.share##

Articole similare

1-10 of 277

Puteți, de asemenea, începeți o căutare avansată de similaritate pentru acest articol.